Study found different sterilization technique could prevent this complication
WebMD News from HealthDay
By Robert Preidt
HealthDay Reporter
TUESDAY, Oct. 7, 2014 (HealthDay News) -- An E. coli outbreak at an Illinois hospital was caused by endoscopes that had bacterial contamination despite being disinfected in the recommended way, a new study says.
The outbreak occurred among patients who underwent procedures with duodenoscopes, which are specialized endoscopes used to diagnose and treat problems in the bile and pancreatic ducts. These are not the same type of endoscopes used for routine endoscopy of the upper digestive tract or for colonoscopies, according to the researchers.
"The complicated design of duodenoscopes makes cleaning difficult. It appears that these devices have the potential to remain contaminated with pathogenic bacteria even after recommended reprocessing is performed," Dr. Lauren Epstein, of the U.S. Centers for Disease Control and Prevention, and colleagues wrote in the study.
"Facilities should be aware of the potential for transmission of antimicrobial-resistant organisms via this route and should conduct regular reviews of their duodenoscope reprocessing procedures to ensure optimal manual cleaning and disinfection," they concluded.
The study appears in the Oct. 8 issue of the Journal of the American Medical Association.
A number of patients were infected with, or exposed to, a highly drug-resistant strain of E. coli through duodenoscope procedures at the teaching hospital between January and December 2013, according to the study.
Investigators found that the hospital followed all manufacturer-recommended duodenoscope reprocessing (such as cleaning) instructions. After the hospital changed to a gas sterilization method, there were no more cases of this type of E. coli among patients who underwent duodenoscope procedures.
Health care providers should report and publish all cases of infectious diseases linked to endoscopy, particularly if recommended reprocessing procedures were followed, in order to determine whether this study "is the tip of the iceberg or an isolated occurrence," William Rutala and Dr. David Weber, from University of North Carolina Health Care, wrote in an accompanying editorial.
"If the former, then revision of the endoscope reprocessing guidelines will be necessary to ensure patient safety," they said.
source : Despite Proper Cleaning, Endoscopes May Pass on E. coli